Chapter 3 Installing the Software and Powering On the Instrument
Troubleshooting
34 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Condition: CCD, Filter Wheel, or Shutter Test Failures
1. Power off the instrument, wait 30 seconds, then power on the
instrument.
2. Perform the Instrument Function Test (see page 30).
3. If the test fails again, perform the ROI calibration as explained
in “Performing the Regions of Interest (ROI) Calibration” on
page 35.
4. Determine if the ROI image appears in the ROI Inspector.
Yes Continue with the installation.
No Contact Applied Biosystems technical support (see
page xiii) or your service representative.
Condition: Lamp Test Failure
1. Power off the instrument, wait 30 seconds, then power on the
instrument.
2. Perform the Instrument Function Test (see page 30).
3. If the test fails again, select Instrument
Lamp
Status/Replacement.
4. If the Lamp Status/Replacement dialog box reports the lamp
status as Failed, replace the halogen lamp (see “Replacing the
Halogen Lamp” on page 118).
5. Perform the Instrument Function Test (see page 30).
6. If the test fails again, contact Applied Biosystems technical
support (see page xiii) or your service representative.
Condition: Thermal Cycler Test Failure
1. Power off the instrument, wait 30 seconds, then power on the
instrument.
2. Perform the Instrument Function Test (see page 30).
3. If the test fails again, contact Applied Biosystems technical
support (see page xiii) or your service representative.
Troubleshooting – Function Tests (continued)
Lamp
status

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)

Related product manuals