Chapter 5 Performing the Background Calibration and Optical Calibration
Overview
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 51
Notes
5
The software then uses the calibration file during subsequent runs to remove the
background fluorescence from the run data.
Background
Fluorescence
Fluorescence data collected by the Applied Biosystems 7300/7500/7500 Fast Real-Time
PCR System includes a fluorescent signal inherent to the system, commonly referred to
as background fluorescence. Background fluorescence is a composite signal found in all
spectral data. This signal consists of fluorescence from several sources, including:
Background electronic signal
Contaminants in the sample block
The plastic consumable (plates and caps)
Guidelines for
Calibration
For a new instrument:
Always start with a new calibration kit.
Make sure the centrifuge you use is clean. Before centrifuging, wipe down the
bucket using a tissue.
Handle the calibration plates with care to prevent contamination. Do not place
plates on a lab bench, which may contaminate the plate. Always put calibration
plates back into their original bags.

Table of Contents

Other manuals for Applied Biosystems 7500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 and is the answer not in the manual?

Applied Biosystems 7500 Specifications

General IconGeneral
Detection MethodFluorescence
Sample Capacity96-well plate
Thermal CyclerYes
Temperature Uniformity±0.5°C
TypeReal-Time PCR System
Thermal Accuracy±0.25°C
Detection Channels5
Weight45 kg
Power Requirements100-240 V, 50/60 Hz
Dynamic Range9 logs
Reaction Volume Range10-100 µL
Temperature Range4°C to 100°C
Thermal Cycler Block96-well
SoftwareSDS Software
DetectionMultiplex (up to 5 colors)