Chapter 4 Install the Standalone Layout
Analyze the RNase P Experiment
Applied Biosystems StepOne
and StepOnePlus
Real-Time PCR Systems
Installation, Networking, and Maintenance Guide
70
Notes
5. Review the standard curve:
a. Select the Standard Curve tab.
b. Click the upper-right corner of the Plate Layout to select all wells.
c. Confirm that the R2 value is greater than or equal to 0.990.
If the R2 value is less than 0.990, repeat the RNase P experiment using a different
RNase P plate. If the problem persists, contact Applied Biosystems as explained in
“How to Obtain Support” on page x.
6. Click Save Status, then click Ye s when prompted.
7. Touch the StepOne
instrument touchscreen to awaken it, then touch Pass in the
RNase P wizard.
5a
5c
5b
6