External Measurements > Temperature > Calibration > Temperature
Calibration temperature at the time of calibration.*
Selection of digit position
Confirmation of setup and saving to memory
Refference temperature of known temperature sensor at the time of calibration
External Measurements > Temperature > Calibration > Star Now
Start of the calibration.
Confirmation of setup and saving to memory
Calibration temperature will not change
New calibration temperature changed
3.6.2. External Measurements > Pressure
Setup of the external measurement of pressure.
Enter this part of the menu
External Measurements > Pressure > Measurement
Confirmation of setup and saving to memory
When set ON, the value of pressure is shown
amongst totalizers (XPress)*
It is not possible to use simultaneously Pressure Module a I-Out
External Measurements > Pressure > Current Loop
4-20mA external pressure sensor settings.*
Possible range: 4 – 20mA, Pressure – according to sensor
Selection of digit position
Confirmation of setup and saving to memory
Any external pressure sensor working on passive 4-20mA. i.e. WIKA A-10 (P#1105VX3J)