Operating Instructions Xenotest
®
150 S+
tight connection and possible leakage of water hoses,
condition of the water filter,
filling level of the instrument-integral water reservoir or of the supply reservoir
for treated water,
condition of connector and power supply cable.
6.3 Installation and removing the xenon lamp
WARNING - Electric shock!
Contact with voltage-carrying components may cause a lethal
electric shock. The irradiator ignition system is under high voltage.
Prior to removing / installing the xenon lamps, disconnect the in-
strument from the power supply system!
Switch the instrument off using the ON/OFF switch.
Disconnect the mains plug from the socket and protect it from acci-
dental reconnection.
Check to see if the instrument is de-energized.
CAUTION - Risk of burns!
The components of the irradiation unit reach extremely high tem-
peratures; contact with skin may cause burns.
Prior to replacing the xenon lamp, allow the irradiation unit com-
ponents to cool down sufficiently!
NOTE - Handling the xenon lamp:
Residues from finger contact may burn into the xenon lamp glass
tube and cause uneven irradiation.
Grasp the xenon lamp only at the lamp socket.
Wear clean cotton gloves for installation.
6 Start-up

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Atlas XENOTEST 150 S+ and is the answer not in the manual?

Atlas XENOTEST 150 S+ Specifications

General IconGeneral
BrandAtlas
ModelXENOTEST 150 S+
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals