56
Operating Instructions Xenotest
®
150 S+
Power supply:
Nominal voltage 230 V ± 10 %, 50/60 Hz
Mains connection (1P, N, PE) or (2P, PE) CEE (32A, 3 poles, 6h)
Nominal output approx 5 kVA max
Xenon lamp voltage 2200 VA max
Xenon lamp current 16 A max
Dimensions and weights:
Dimensions 900 mm x 780 mm x 1800 mm (Width x Depth x Height)
Weight approx 280 kg
Available sample surface:
Non-turning mode approx 500 cm²
Turning mode approx 950 cm²
Noise:
Noise level 65 db(A)
Cooling air requirement:
Lamp unit 200 m³/h max
Test chamber 100 m³/h max
Water consumption:
Treated water / misting 0.7 l/min max
Treated water / humidification 0.033 l/min max
Ambient conditions:
Room temperature 18 °C to 25 °C
Relative humidity 50 % (+/- 10 %)
Transport and storage conditions:
Temperature - 25 °C to 70 °C
Relative humidity 75 % max
10 Technical data