APPENDIX
Technical Data
93
J
Temperature
Incubator air temperature
control mode
Time proportional control mode
Incubator temperature
setting range
23.0-38.0°C (in 0.1°C increments)
Incubator air temperature
display range
20.0-42.0°C (accuracy: ±1.0°C)
Skin temperature display
range
30.0-42.0°C (accuracy: ±0.3°C)
Heater output 0-100 (indicated in 10 levels)
Warm-up time Approx. 40 min at ambient temperature of 25°C
Alarms Excessive temperature, Set temperature, Fan, Skin temperature probe, Sys-
tem
J
Oxygen concentration
Type of oxygen sensor Galvanic cell type
Oxygen concentration
display range
15-105% Accuracy: 15-25%: ±2%O
2
25-100%: ±3%O
2
Setting range of oxygen
level alarms
Upper limit: 22-99%O
2
(in 1% increments)
Lower limit: 19-96%O
2
(in 1% increments)
Response time for
oxygen concentration
display
30 sec (90% response)
Stability of oxygen
concentration
measurements and
calibration intervals
24 hours
Calibration 21%
Useful life of oxygen
sensor
1 year after start of use
Alarms Oxygen sensor and upper/lower alarm limits
J
Oxygen supply
Maximum oxygen
concentration
≥ 60% (at O
2
flow rate of 10L/min)
Time to maximum blood
concentration
≤ 30 min (from 21-55% at O
2
flow rate of 10L/min)