Environment
CO
2
concentration in the
hood
≤0.5% when a 4% mixture of CO
2
in the air is delivered at 750mL/min at a
point 10cm above the center of mattress
J
Pulse oximeter
1, 5, 6, 7,10
(Masimo pulse oximeter)
SpO
2
Display range: 1-100%
Measurement accuracy: During no motion
2
±3 digits (70-100%)
During motion
3
±3 digits (70-100%)
During low perfusion
4
±3 digits (70-100%)
Update cycle: 1 second
Alarm setting range: Upper limit OFF, 50-99% (in 1% increments)
8
Lower limit OFF, 45-95% (in 1% increments)
8
Pulse rate Display range: 25-240bpm
Measurement accuracy: During no motion
2
±3 digits (25-240bpm)
During motion
3
±5 digits (25-240bpm)
During low perfusion
4
±3 digits (25-240bpm)
Update cycle: 1 second
Alarm setting range: Upper limit OFF, 80-235bpm (in 5bpm increments)
8
Lower limit OFF, 35-180bpm (in 5bpm increments)
8
Sensor
characteristic
Wavelength: 660nm (red light), 905nm (infrared light)
Brightness: ≤15mW
Non-sterile, latex-free
Mode setting Averaging time: 2, 4, 8, 10, 12, 14, 16 sec (selectable)
8
(Factory default: 8 sec)
Sensitivity setting: APOD, Normal (normal sensitivity), Max (high sensitivity)
8 9
(Factory default: Normal)
Available
patient cables
RD rainbow SET MD20-05, 1.5m
Available
disposable
sensor
RD SET Neo
RD SET Inf
RD SET NeoPt
1 This device is calibrated to functional oxygen saturation.
2 The pulse oximeters have been validated for no motion accuracy in human blood studies on healthy
adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies in the
range of 70-100% SpO
2
against a laboratory CO-Oximeter and ECG monitor. This variation equals plus
or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the popula-
tion.
3 The pulse oximeters have been validated for motion accuracy in human blood studies on healthy adult
male and female volunteers with light to dark skin pigmentation in induced hypoxia studies while per-
forming rubbing and tapping motions, at 2-4Hz at an amplitude of 1-2cm and a non-repetitive motion be-
tween 1-5Hz at an amplitude of 2-3cm in induced hypoxia studies in the range of 70-100% SpO
2
against a
laboratory CO-Oximeter and ECG monitor. This variation equals plus or minus one standard deviation.
Plus or minus one standard deviation encompasses 68% of the population.