APPENDIX
Technical Data
95
4 In a laboratory test, in the range of 70-100% SpO
2
, accuracy in low perfusion conditions were checked
against the Biotec Index2 Simulator and the Masimo Simulator, both of whose signal strength is 0.02%
or more and permeability 5% or more. This variation equals plus or minus one standard deviation which
encompasses 68% of the population.
5 The materials accessible to the patient and the user comply with ISO 10993-1.
6 All the sensors and patient cables that can be used with this device have been tested and verified by us-
ing the Masimo/MX series pulse oximeter technology.
7 Since the measurements taken by the pulse oximetry equipment are statistically distributed, only about
two-thirds of them fall in the accuracy range (ARMS) of values obtained with a CO-oximeter.
8 Even if the power supply is interrupted due to power failure, disconnection of the power plug or some
other cause, the preselected settings will be retained in the memory. When the power supply is re-
sumed, the unit will start operation with the last selected settings.
9 The Max sensitivity setting will not be stored in the unit. After a power off and on cycle, the sensitivity
will change from the “Max” to “Normal” (see “9-1-6. Setting the Sensitivity Mode”).
10 Alarm delays should not exceed 30 seconds other than as specified in this manual.
J
Pulse oximeter
1, 4, 5, 7
(Nellcor pulse oximeter)
SpO
2
Display range: 1-100%
Measurement accuracy: During no motion
2
±3 digits (70-100%)
During low perfusion
3
±3 digits (70-100%)
Update cycle: 2 second
Alarm setting range: Upper limit OFF, 50-99% (in 1% increments)
6
Lower limit OFF, 45-95% (in 1% increments)
6
Pulse rate Display range: 25-240 bpm
Measurement accuracy: During no motion
2
±3 digits (25-240bpm)
During low perfusion
3
±3 digits (25-240bpm)
Update cycle: 2 second
Alarm setting range: Upper limit OFF, 80-235bpm (in 5bpm increments)
6
Lower limit OFF, 35-180bpm (in 5bpm increments)
6
Sensor
characteristic
Wavelength: 660nm (red light), 900nm (infrared light)
Brightness: <15mW
EOG sterilization, latex-free
Mode setting Response mode: Normal, Fast
6
(Factory default: Normal)
Available
pulse oxim-
etry cable
DOC-4, 1.2m
Available
disposable
sensor
SoftCare SC-PR
OXISENSOR III N-25
OXISENSOR III I-20