Miscellaneous Menu
Misc
Settling
[Read Only]
The time the magnetic circuit takes to settle. An engineering diagnostic.
Language
[Language]
U
Allows you to change the current language. English is the default setting.
The following languages are supported:
German (Deutsch), Czech (Cestina), Spanish (Espanol), French (Francais), Russian (России),
Italian (Italiano).
Date
[Date DMY]
[Read Only]
S
Real-time calendar. The day, month and year must be reprogrammed after the battery is replaced or
powered off.
Time
[Time]
S
Real-time clock. The hour, minute and second must be reprogrammed after the battery is replaced or
powered off.
EEPROM
[EEPROM]
A
Format the EEPROM to erase all log files. Totalizers and configuration remain unaffected during
aformat.
Battery
[Battery]
S
Saves Totalizers to nonvolatile memory in preparation for battery replacement.
Restart
[Restart]
S
Provides the ability to reset the meter electronics.
HDD Free
[HDD Free]
(Read only)
S
Indicates the amount of free flash memory space.
Polarity
[Polar V]
(Read only)
Measured electrode polarizing voltage (for service purposes).
Datalogger
[DataLog]
The logging period can be adjusted to the following values:
1 min / 15 min / 1 h / 6 h / 12 h / 24 h. See the Datalogging User Manual for more information.
LCD Test
[LCD Test]
After pressing E, all display segments appear for about 2 seconds.