CONNECTING AN ORION RTR® ENDPOINT TO THE M5000 METER
OTE: N Connect the endpoint as described. If there is a reading on the M5000, program the endpoint to match that reading
or reset the totalizer. The endpoint requires programming per the endpoint user manual.
Wiring
To connect the RTR endpoint to Output #1 on the M5000, connect the red wire to the positive (+) terminal and the black and
green wire to the negative (–) terminal.
To connect the RTR endpoint to Output #2 on the M5000, connect the red wire to the positive (+) terminal and the black and
green wire to the negative (–) terminal.
Programming
To program the M5000 meter for the endpoint to Output #1 (forward flow):
1. Navigate to IN/OUT > Simulat > Outputs > Puls/unit.
2. Use the arrows to change the values, then press EXIT/SAVE.
3. Repeat steps 1 and 2 for Width, Out 1 Func and Out 1 Type.
OTE: N For Output #2 (forward flow), use Out 2 Func and Out 2 Type instead.