Programming the Pump 51
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
The SIGMA Spectrum Infusion System is not intended to replace
clinician patient observation.
The Pump was not designed nor is it intended to detect infiltrations or
extravasation
WARNING Confirm Drug Library.
Master Drug Library Administrators (MDLAs) should verify the
correct Drug Library is installed when deploying the Drug Library to
Pumps.
Master Drug Library Administrators (MDLAs) should verify the
Drug Library transfer is successful after deployment.
Users should verify the correct Drug Library is installed on the
deployed Pumps.
Before implementation, clinical users at each facility must
thoroughly test and validate their Drug Library per their facility’s
procedure to ensure configuration and workflow reflect clinical
practice.
CAUTION Confirm Audio Operation.
Listen for beeps when pressing keys. If sound is not heard,
discontinue use of the Pump and refer servicing to qualified service
personnel at your facility or return the Pump to Baxter for service.
CAUTION Confirm Display Operation.
Regularly observe the Pump’s display. If display abnormalities are
observed, discontinue use of the Pump and refer servicing to
qualified service personnel at your facility or return the Pump to
Baxter for service.
To program the Pump using DERS:
1. Press ON/OFF to power on the Pump.
2. The New Patient screen may display.
The Pump retains previously programmed infusion values up to 24 hours after power off. The
New Patient screen will display at power on to allow use of the previously programmed values
or allow a new infusion to be programmed.
3. At the New Patient screen:
Press the yes soft key to clear the previously programmed infusion values and allow
programming a new infusion.
NOTE: If it is the same patient, but a new drug is required, press the yes soft key. As a
result, the Total given will be reset to zero.