Error
Code
Error Message Possible Cause Error Resolution
E041
Container store:
destination location
not empty detected
for transfer
The specified location
already contains a
plate.
1. Select Reset Error [F5].
2. Open the ReadA incubator
and identify the location the
gripper is trying to transfer a
plate into.
a. If there is a plate in the
location, close the door and
select Start [F2] to restart
the module.
The plate will be sent to the
ReadA outfeed and then to
the Track. Once it is
re-scanned, it will be sent
back to the ReadA.
b. If there is no plate in the
location, check the plate
detection sensor and clean
or adjust it if necessary.
Select Start [F2] to restart
the module.