EasyManuals Logo

Beckman Coulter COULTER ACT 5diff User Manual

Beckman Coulter COULTER ACT 5diff
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #165 background imageLoading...
Page #165 background image
PN 177196BB
8-13
SYSTEM DIAGNOSTICS AND MAINTENANCE
RECOGNIZING NORMAL INSTRUMENT OPERATION - CYCLE DESCRIPTION
8
Double Rinsing the Sampling Probe
1. The horizontal traverse assembly moves the
sampling probe over the RINSE bath.
2. A double rinsing (interior and exterior) of the
probe removes residual whole-blood sample
from inside the probe.
a. In the CBC/DIFF mode, 5 µL is discarded
in the RINSE bath.
b. In the CBC mode, 7 µL is discarded in the
RINSE bath.
Aspirating from the First Dilution
1. The horizontal traverse assembly moves the
sampling probe over the DIL 1/HGB bath.
2. The vertical traverse assembly moves the probe
downward into the bath.
3. 42.5 µL of the 1:170 first dilution is aspirated
into the sampling probe.
DIL 1
HGB
DIFF RBC BASO
WBC
RINSE
DIL 1
HGB
DIFF RBC BASO
WBC
RINSE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter COULTER ACT 5diff and is the answer not in the manual?

Beckman Coulter COULTER ACT 5diff Specifications

General IconGeneral
BrandBeckman Coulter
ModelCOULTER ACT 5diff
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals