PN 177196BB
3-11
SETUP OPTIONS
SYSTEM SETUP OPTIONS
3
1) The Enable box allows you to include any of the following:
Note: With the exception of
Raw Data, all selections are enabled by default.
a)
Ranges. Normal ranges for each parameter reported.
b)
Raw Data. The raw values collected during analysis.
Note: This data can be useful as a troubleshooting tool, but do not enable
this option unless instructed by your Beckman Coulter Representative.
c)
Microscopic Examination Area. A labeled area where results of the
microscopic examination can be added.
d) Messages. Interpretive messages, such as Anemia when the Hgb < Hgb LL.
For more information, refer to Interpretive Messages under Heading 9.8
FLAGS AND MESSAGES GENERATED BY THE INSTRUMENT, in the
Online Help System or the Instructions for Use manual.
e)
Histogram and DiffPlot. The histograms and DiffPlots generated as part of the
sample analysis.
2) The Display and Print
Detailed Flags option allows you to choose between
printing (and displaying) detailed flags or suspect flags for a parameter.
a)
Detailed Flags enabled is the default.
b) As a general rule the detailed flag is more parameter specific, the suspect
flag more generic.
c) To see the differences, refer to Suspect Flag Format under Heading 9.8
FLAGS AND MESSAGES GENERATED BY THE INSTRUMENT, in the
Online Help System or the Instructions for Use manual.
3) The Hematology Parameters Printed box allows you to select the whole blood
parameters to include on the report.
4) Have the trainee find the procedure, Patient Report Setup, in Appendix
A.8 of the
Online Help System or the Instructions for Use manual, and make the relevant
selections for their laboratory.
6. Click the
Cycle Options tab.
a. The Cycle Options screen allows you to define five cycle options. Explain each,
having the trainee change settings if needed.
Note: To change settings, (1) click , (2) edit the setting, then (3) click to
accept the changes.
b. Auto-Clean
1) The Auto-Clean field allows you to enter the number of analyses between
automatic Auto-Clean cycles.
2) In the Auto-Clean cycle, Rinse reagent is dispensed to the bath assemblies to
clean the bath assemblies and the apertures, and then flushed out with Diluent.
3) You can select an interval from 1 to 120 analyses between Auto-Clean cycles.
4) The default number is 100.
c. Daily Workload
1) The Daily Workload fields allow you to enter the number of CBC and
CBC/DIFF samples typically ran in a day.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter COULTER ACT 5diff and is the answer not in the manual?

Beckman Coulter COULTER ACT 5diff Specifications

General IconGeneral
BrandBeckman Coulter
ModelCOULTER ACT 5diff
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals