EasyManua.ls Logo

Bender ATICS-2-63A-ISO - Functional Safety; Product Life-Cycle Management; Application Examples; Application Example Operating Theatre

Bender ATICS-2-63A-ISO
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14 ATICS-2-ISO_D00046_04_M_XXEN / 03.2021
System description
3.2 Functional safety
Functional safety according to IEC 61508 guarantees safety from risks due to malfunction when used
properly. The ATICS® is suitable for use in safety-related systems according to SIL2. SIL stands for "Safety
Integrity Level".
3.2.1 Product life-cycle management
Safety must be guaranteed over the entire life cycle, from the time it is designed, developed, manufac-
tured, commissioned, maintained to the time it is taken out of service. Responsibility during the life cy-
cle:
For detailed information refer to:
Chapter „Commissioning, settings and testing“ on p. 45.
chapter „Periodic verification and service“ on p. 89.
3.3 Application examples
3.3.1 Application example operating theatre
ATICS-2-63A-ISO: Changeover be-
tween the preferred and redun-
dant line while monitoring the
medical IT system with transform-
er load and temperature monitor-
ing
IR426-D47: Monitoring of the op-
erating theatre light IT system
(optional)
MK2430/MK800/TM800/CP9xx:
Alarm at at least two points with
independent power supplies for
functional safety
Production Commissioning RecyclingDecommissioningMaintenance
Bender plant operator Bender
U
1
U
2
U
3
I
3
I
L
R
ISO
T
ATiCS-2-63A-ISO
ϑ
2 2
BSV
BMS
MK
IR426-D47
TM800
Second
supply line
Preferred
supply line
Nurse desk
Operating room

Table of Contents