EasyManua.ls Logo

Bender ATICS-2-63A-ISO - Page 88

Bender ATICS-2-63A-ISO
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
88 ATICS-2-ISO_D00046_04_M_XXEN / 03.2021
Troubleshooting
Why does the ATICS® not display the set time?
The time of the ATICS® is set by the BMS bus master. The time for the ATICS® only is required in the rare
case when no BMS bus is used. In this case, ATICS® displays the fault message "No master".
How do I eliminate the message "No master"?
ATICS® is designed for applications where functional safety is of top priority. Therefore, this message is
necessary. For applications where functional safety only plays a subordinate role, master monitoring
can be disabled in the menu "Settings > Interface > Failure monitoring".
ATICS® does not switch over. In the last line on the display it indicates: "t(0): xx s".
ATICS® is waiting for the time delay t(0) to elapse and displays the remaining time in seconds. Only then
can the ATICS® switch over. The time delay can be due to the following reasons:
ATICS® is just switching. The set time delay in switch position "0" (dead time) is several sec-
onds (see „Settings menu 1: Changeover“ on p. 67)
ATICS® has just been supplied with voltage. After the time delay has elapsed, ATICS® is ready
for operation.
Within a short period of time, ATICS® carried out more than 12 changeover operations. The
time delay is required to cool the system.
Why does writing back reports/backups lead to error messages?
ATICS® – like many Bender devices – can be monitored and configured by means of a Bender gateway
CP700, COM46x… etc. The gateways are operated via a web user interface that is displayed using an
Internet browser.
One of the functions of these gateways is saving reports. A report is used to store the settings and
measured values of a bus device. By creating a backup, data loss (e.g. by accidentally deleting settings)
can be avoided. The backup can also be imported to another device of the same type. This may be nec-
essary if a device is replaced or if several devices are to be configured for similar tasks.
ATICS® is continuously further developed and maintained. This also affects the content of the report file.
If you try to save a report that was generated with an older ATICS® (version older than 1.20) in an ATICS®
with a current firmware, this can lead to error messages. These error messages are displayed on the
web user interface.
Repeat the saving process of the report in ATICS® several times. This way, the majority of the
parameters can still be transferred. The remaining error messages indicate the parameters
that have to be changed manually.
To conclude, check all settings saved in the ATICS®.
Save these settings in a new report file.

Table of Contents