Installation and connection
34
MK2430_D00129_02_M_XXEN/09.2019
Example 3: Intensive care unit with two IT systems and four rooms
Device Parameter Address settings
First changeover and monitoring module
ATICS "A" Bus address 3
EDS151 "A" Bus address 4
Second changeover and monitoring module
ATICS "B" Bus address 5
EDS151 "B" Bus address 6
Alarm indicator and test combinations
First MK2430…
Bus address 1
Test address 3, 5
Alarm address 2, 3, 4, 5, 6, 7, 8
Second
MK2430…
Bus address 2
Test address 3, 5
Alarm address 1, 3, 4, 5, 6, 7, 8
Third
MK2430…
Bus address 7
Test address 3, 5
Alarm address 1, 2, 3, 4, 5, 6, 8
Fourth
MK2430…
Bus address 8
Test address 3, 5
Alarm address 1, 2, 3, 4, 5, 6, 7
„A“
ATICS
EDS151
„B“
ATICS
EDS151