100
Clinician's Guide
Guidance and Manufacturer’s Declaration—
Electromagnetic Immunity for All Equipment and Systems
The Vector Elite System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Vector Elite System should assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment—
Guidance
Voltage
dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip
in U
T
) for 5 sec
<5% U
T
(>95%
dip in U
T
) for 0.5
cycle
40% U
T
(60%
dip in U
T
) for 5
cycles
70% U
T
(30%
dip in U
T
) for 25
cycles
<5% U
T
(>95%
dip in U
T
) for 5
sec
Mains power quality should be that
of a typical hospital environment. If
the user of the Vector Elite System
requires continued operation during
power mains interruptions, it is
recommended that the equipment
be powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 Aa/m 30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The Vector Elite System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Vector Elite System should assure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment—Guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the Vector Elite System (trolley, hand-held and
PC), than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.