4
Clinician's Guide
Ensure that the handheld is operating and communicating with the appropriate computer before
starting a training session.
Warnings
Incorrect patient data entry or incorrect patient retrieval may cause the patient to fall and/or
serious injury.
Do not place a patient that weighs less than 30 lbs. (13 kgs.) or more than 500 lbs. (227 kgs.) in
the Vector Gait & Safety System.
Do not exceed recommended weight limits or otherwise misuse the equipment. The equipment
can dislocate and fall causing serious injury.
Do not lift the patient too high or have the patient ascend/descend high stairs where they might be
injured from the trolley.
Do not touch the motorized Vector Elite trolley during a training session. The trolley contains
moving parts that could cause injury if touched.
In an emergency such as an electrical outage, stop exercise immediately to avoid injury and turn
off the power switch located on the power station.
Do not attempt to service or repair the Vector Elite system or electrocution may occur.
Modifying the Vector Elite system equipment voids the warranty.
Do not use the Vector Elite system if maintenance is required; serious injury may occur.
In the event of power failure, carefully remove patient from the harness and step away from the
equipment.
Do not use equipment if the Vector Elite trolley has become wet (e.g., from sprinklers) as there is
a risk of electrocution. Servicing is required before the equipment may be used.
To avoid risk of electric shock, this equipment must only be connected to a supply mains with
protective earth.
The Vector Elite system is not intended for patient transport. Using the system as a transport could
cause serious patient injury.
Manually moving the trolley along the track by pulling on the Support Rope, could damage the
Vector Elite system.
The Vector Gait & Safety System may only be operated by a trained health care professional.
The use of accessories not specified by the manufacturer may result in injury and/or increased
emissions, or decreased immunity of the equipment, affecting its performance.
There is a possibility of electromagnetic interference with other devices. When there is suspicion
of possible EMC interference to the Vector Gait & Safety System, it is recommended to shut down
the electrical devices nearby, one by one, to locate the possible interference source.
Due to potential confusion about which computer is being used with which patient, a clinician
should not operate multiple computers at the same time.
The Vector PC and Remote Control network configurations should not be modified without
direction from Bioness Support, including changing network connection or connection to the