ARRHYTH: Sets up ON/OFF to indicate detection of diagnosis (Asys, VTAC/VFIB and VTAC).
OFF: Do not perform arrhythmia diagnosis.
LETHAL: Performs the detection of Asys, VTAC/VFIB, and VTAC at the selected lead
FULL: Performs the detection of all 13 arrhythmia.
The Analysis algorithm simultaneously uses leads I, II, III, and the V lead for ECG and arrhythmia
analysis.
PVC
ANALYSIS:
ON
ARRHYTH
LEVEL
PACE :
OFF
ARRHYTHM :
OFF
PREV
MENU
MAIN
MENU
ST
SETTING
ECG
FILTER :
MONITOR
PVC
ANALYSIS:
ON
ARRHYTH
LEVEL
PACE :
OFF
ARRHYTHM :
LETHAL
PREV
MENU
MAIN
MENU
ST
SETTING
ECG
FILTER :
MONITOR
PVC
ANALYSIS:
ON
ARRHYTH
LEVEL
PACE :
OFF
ARRHYTHM :
FULL
PREV
MENU
MAIN
MENU
ST
SETTING
ECG
FILTER :
MONITOR
ACC VENT
Adult— Accelerated ventricular occurs when six or more ventricular beats are detected
with an average heart rate for the ventricular beat between 50 and 100 beats per minute.
0-2 years—Occurs when six or more ventricular beats are detected with an average heart rate for
the ventricular beat between 60 and 160 beats per minute.
3-10 years—Occurs when six or more ventricular beats are detected with an average heart rate for
the ventricular beat between 60 and 140 beats per minute.
11-13 years—Occurs when six or more ventricular beats are detected with an average heart rate for
the ventricular beat between 60 and 130 beats per minute.
ASYSTOLE
Ventricular asystole occurs whenever the displayed heart rate drops to zero.