Rev 3.1 10. EtCO2 169
BM5 (CS,CX) User’s Manual
10.3 EtCO2 Parameter Setting Menu
ALARM : A menu to set the alarm limit and sound.
SWEEP SPEED: EtCO2 sweep speed is changeable to 6.25, 12.5, 25mm/s.
WAVEFORM SCALE: A menu to set the screen scale of measured waveform.
APNEA DETECT : Deciding function of activating Apnea Alarm.
SETTINGS: A menu to handle the information of EtCO2 signal
ZERO : When a new sensor is connected, the message “ZERO REQUIRED” is displayed.
ZEROSETTINGS
APNEA
DETECT:
OFF
ALARM
SWEEP
SPEED:
6.25mm/s
WAVEFORM
SCALE:
40mmHg
MAIN
MENU
ALARM LIMIT(Upper/lower limit value of alarm)
Upper/lower limit value of alarm differs depending on the position of measurement.
The basic setting range of alarm setting value for EtCO2, FiCO2, awRR, APNEA.
ZEROSETTINGS
APNEA
DETECT:
OFF
ALARM
SWEEP
SPEED:
6.25mm/s
WAVEFORM
SCALE:
40mmHg
MAIN
MENU
ALARM
LIMIT
ALARM
SOUND:
OFF
PREV
MENU
MAIN
MENU