BM5 (CS,CX) User’s Manual
Cleaning Applied Parts
Do not permit any liquid to enter the monitor case and avoid pouring it on the monitor while
cleaning. Do not allow water or cleaning solution to enter the connectors of jack cover.
Recommended cleaning agents:
Alcohol (Ethanol 70%, Iosopropanol 70%, Window cleaner)
Ammonias (Dilution of ammonia <3%, Window cleaner)
Tensides (dishwasher detergents) (Edisonite schnellreiniger
®
, Alconox
®
)
Cables and Leadwires
Do not use acetone or keytone solvents for cleaning; do not use an autoclave or steam cleaner.
Cables and leadwires can be cleaned with a warm, damp cloth and mild soap, or isopropyl alcohol
wipes. For more intensive disinfecting (near sterile) Ethylene Oxide (ETO) is acceptable but will
reduce the useful lifetime of the cable or leadwire.
The decision to sterilize must be made per your institution’s requirements with an awareness of
the effect on the integrity of the cable or leadwire.
The Equipment needs safety inspection once a year. Please refer to user’s guide or service
manual for the examine objects.
Please check carefully both frame and sensor, after cleaning the Equipment, Do not use the
equipment that is worn out or damaged.
At least once a month, clean and wipe off the frame by using the soft cloth after wetting it with water
and alcohol. Do not use lacquer, thinner, ethylene, and oxidizer which may leads damage to the
equipment.