5.2 ECG Data Window
ECG Wave Display is always on when the cable is connected.
The heart rate is calculated by a moving average. The monitor detects 8 consecutive beats,
averages the R-R intervals of the latest 8 beats and uses this average to calculate the current
heart rate. When a new beat is detected, the heart rate is recalculated using the latest 8beats. The
heart rate display is updated every 3 seconds.
Heart rate meter updates a new heart rate for a step increase or decrease in 10 seconds maximum.
When ventricular tachycardia is detected, the alarm set in 5 seconds maximum.
Check that the delay time of the output signal (alarm trigger 80ms maximum) is within the range of
the connected equipment.
Heart Rate Alarm Limit:
Decides the QRS limits, and
gives an alarm if a value is over
the limits.
QRS: Detects QRS, and
flashes when QRS is
detected.
Heart Rate:
Displays heart rate per minute.
Pace Detector
Indicators:
Detects and displays
the patient’s pace
maker and flashes.