Rev. 3.1 8. NIBP 145
BM5 (CS,CX) User’s Manual
INTERVAL
This menu is used for selecting intervals when measures the blood pressure automatically.
Select a target interval from 1min, 2, 3, 4, 5, 10, 15, 20, 30, 1hour, 2, 4, 8.
INTERVAL:
OFF
INFLATION:
170mmHg
UNIT
SELECT:
mmHg
ALARM
NIBP
STAT:
OFF
CUFF
SIZE:
ADT
MAIN
MENU
> OFF 15MINS
1MIN 20MINS
2MINS 30MINS
3MINS 1HR
4MINS 2HRS
5MINS 4HRS
10MINS 8HRS
INTERVAL:
OFF
MAIN
MENU
Periodically check patient limb circulation distal to the cuff. Check frequently when using auto
NBP in 1 and 2 minute intervals. Intervals below 10 minutes are not recommended for extended
periods of time.
INFLATION
It is a function for set the maximum initial inflation pressure value.
The range of initial inflation pressure value of BM5 (CS, CX) is as follows.
ADT/PED : Numeric value is 80, 90, 100, 110, ~ 230, and 240.
Numeric value is 60, 70, 80, 90, 100, 110, and 120.
INTERVAL:
OFF
INFLATION:
170mmHg
UNIT
SELECT:
mmHg
ALARM
CUFF
SIZE:
ADT
MAIN
MENU