Rev. 3.1 15. TROUBLE SHOOTING
203
BM5 (CS,CX) Users Manual
15.2 SpO2 malfunction
Connectors of the equipments are in bad condition?
The extension cables Yes
are disconnected
No
The Finger probe is Yes
in bad condition?
No
Repair the ECG B/D
15.3 Temp malfunction
The probe is Yes
in bad condition?
No
Repair the Temp B/D
Replace extension cables
Replace the finger probe
Replace the probe