Rev 3.1 10. EtCO2 175
BM5 (CS,CX) Users Manual
MODULE INFO
This is information for handling the EtCO2 module.
MODULE
SETUP
MODULE
INFO
MODULE
RESET
PREV
MENU
MAIN
MENU
RETURN CONTENTS
SENSOR PN 1022054
OEM ID 0X01
SENSOR SN SN6412
MODULE INFORMATION SET
HW REVISION NUM A
TOTAL USE TIME 148830 MIN.
LAST ZERO TIME 0 MIN.
PUMP TOTAL USE TIME 141540 MIN.
PUMP MAX USE TIME 1440000 MIN.
SENSOR PN(part number) : The sensor part number
OEM ID : The id is a 7bit identifier which is set at the factory to a unique value for each OEM.
SENSOR SN : The serial number of the module.
HW REVISION NUM : The hardware version number of the module.
TOTAL USE TIME : Total use time of the module.
LAST ZERO TIME : This is the total time that has elapsed with the sensor in service the last zero.
PUMP TOTAL USE TIME : This is the total time the pump has been on.( LoFlo only)
PUMP MAX USE TIME : This value indicates the maximum rated lifetime of the sampling pump.
(LoFlo only)