Rev 3.1 10. EtCO2 174
BM5 (CS,CX) Users Manual
Mode command [command 80h] and the CO2/O2 Waveform
Mode command [command 90h] ) MUST NOT be active when
this command is used otherwise this command will be ignored
and the setting will remain unchanged.
SLEEP MODE: Sleep mode is used to save power when the host monitor is in
standby mode. There are two sleep modes available for the
Capnostat. Using Sleep Mode 1 maintains the heaters so the
Capnostat is able to run immediately after exiting the sleep
mode. Mode 2 will require the Capnostat to go through its warm
up sequence when exiting this mode and a delay will be
introduced until the system has stabilized.
ZERO GAS TYPE: When performing a zero on room air, this setting should be set
to room air (the default). Only change to nitrogen (N2) when
performing a zero on 100% N2 gas; this is provided for use in a
laboratory environment.
DISABLE SAMPLING PUMP: This setting allows the pump to be forced off. In Normal
Operating Mode, the pump will be turned on when the sampling
cell is connected and no pneumatic system errors are detected.
In Pump Disabled Mode, the pump will remain off in all
circumstances.