Rev. 3.1 1.BASIC
34
BM5 (CS,CX) Users Manual
Equipment Sign
ATTENTION :
Consult accompanying documents
Defibrillator-proof TYPE CF APPLIED PART :
Insulated (floating) applied part suitable for intentional external and internal
application to the patient including direct cardiac application. "Paddles"
outside the box indicate the applied part is defibrillator proof.
Medical Standard Definition :
F-type applied part(floating/insulated) complying with the specified
requirements of IEC 60601-1/UL 2601-1/CSA 601.1
Medical Standards to provide a higher degree of protection against electric
shock tan that provided by type BF applied parts.
Defibrillator-proof TYPE BF APPLIED PART :
Insulated (floating) applied part suitable for intentional external and internal
application to the patient excluding direct cardiac application. "Paddles"
outside the box indicate the applied part is defibrillator proof.
Medical Standard Definition :
F-type applied part (floating/insulated) complying with the specified
requirements of IEC 60601-1/UL 2601-1/CSA 601.1
Medical Standards to provide a higher degree of protection against electric
shock than that provided by type B applied parts.