Make sure both cables and accessories are free of dust or contaminants, and wipe them off with soft
cloth wetted with warm water (40°), and at least once a week, clean them by using the clinical
alcohol.
Do not submerge the accessories under any liquid or detergent. Also, make sure any liquid not to
penetrate into the Equipment or probe.
Disinfecting
Do not mix disinfecting solutions (such as bleach and ammonia) as hazardous gases may result.
Clean equipment before disinfecting.
Recommended disinfecting agents:
Aldehyde based (Cidex
®
activated dialdehyde solution, Gigasept )
Alcohol base (Ethanol 70%, Isopropanol 70%, Spitacid
®
, Streilium fluid
®
, Cutasept
®
, Hospisept
®
,
Tinktur forte, Sagrosept
®
, Kodan
®
)
Do not dispose single use probe to any hazard place, Always think about environmental
contamination.
There is back-up battery on board inside system. When users dispose this battery, Please
waste proper place for environmental protection.
Check the electrodes of batteries before changing them.
· Operate BM5 (CS, CX) with internal electric power supply when unsure of external ground
connection or installation occur.
· Remove the 1st Battery when not using equipment for a while without any damage.
For other applied parts such as temperature sensors, pulse oximetry probes, and NBP cuffs, you
must consult the manufacturer for cleaning, sterilization, or disinfecting methods.