Rev. 3.1 8. NIBP 140
BM5 (CS,CX) Users Manual
The maintenance is performed every 2 years.
Check the following list devise to operates properly and safety at all times.
1. Check for proper cuff size.
2. Check for residual air left in the cuff from a previous measurement.
3. Make sure cuff is not too tight or too loose.
4. Make sure cuff and heart are at same level, otherwise hydrostatic pressure will offset the NIBP
value.
5. Minimize patient movement during measurement.
6. Watch for pulses paradox us.
7. Check for leak in cuff or tubing.
8. Patient may have a weak pulse.