personnel that there is no danger to the patient, the operator, or the environment as a result. In
those instances where there is any element of doubt concerning the safety of connected
devices, the user must contact the manufacturers concerned (or other informed experts) for
proper use. In all cases, safe and proper operation should be verified with the applicable
Manufacturer’s instructions for use, and system standards IEC 60601-1-1/EN 60601-1-1 must
be complied with.
Electrosurgery Unit
Electrosurgical units(ESU) emit a lot of RF interference. If the monitor is used with an
ESU,RF interference may affect the monitor operation.
Locate the monitor as far as possible from the ESU. Locate them on opssite sides of the
operating table, if possible.
Connect the monitor and ESU to different AC outlets located as far as possible from each
other.
When using this monitor with an electrosurgical unit, its return plate and the electrodes for
monitoring must be firmly attached to the patient. If the return plate is not attached
correctly,it may burn the patient’s skin where the electrodes are attached.