SECTION 6 DIAGNOSTICS AND TROUBLESHOOTING
______________________________________________________________________
6-8
Status Messages: Possible Causes: Corrective Actions:
No more Uses for this
Sensor
[ 95 ]
Sensor has been connected
and disconnected too many
times.
Replace the sensor.
Reinsert sensor firmly
[ 96 ]
1. Poor or contaminated
connection between
sensor and PIC.
2. Defective PIC.
3. Defective BISx.
1. Connect/clean connection
between sensor and PIC.
2. Replace the PIC.
3. Replace the BISx.
Sensor used for over
24 hours
[ 109 ]
Sensor was attached to
system for more than 24
hours.
Replace sensor.
Data export error
[ 512 ]
(See Message 3000)
Date and time not
changed
[ 513 ]
A case was in progress when
the time/date was changed.
Change the date/time after the
case has finished.
Trend memory error
[ 516 ]
Trend memory is full or
data has been corrupted.
Disconnect the BISx and
restart monitor. Do not
reconnect the BISx until
directed.
Unrecoverable BISx
error
[ 517 ]
(See Message 1000)
No data available
[ 524 ]
1. No data has been
written to the
BISx/monitor.
2. Defective BISx.
3. Defective monitor.
1. Try again after a case has
been run.
2. Replace BISx.
3. Replace monitor.
PDF creation halted:
USB error
[ 527 ]
1. Removable drive is not
connected properly to
USB(A) port.
2. Drive “write protect” is
locked.
3. Drive is incompatible or
defective.
4. Drive is full.
1. Check connection.
2. Verify “write protect”
switch on the drive is set to
“unlock” position.
3. Replace drive. Contact
your Aspect Technical
Service representative for a
list of compatible drives.
4. Replace drive.