SECTION 6 DIAGNOSTICS AND TROUBLESHOOTING
______________________________________________________________________
6-9
Status Messages: Possible Causes: Corrective Actions:
System error
[ 533 ]
(See Message 2000)
Snapshot corrupted
[ 537 ]
(See Message 6000)
System error
[ 538 ]
(See Message 2000)
Export halted: USB
drive not detected
[ 539 ]
1. Removable drive is not
connected properly to
USB(A) port.
2. Drive is incompatible or
defective.
3. USB port is defective.
1. Check connection.
2. Replace drive.
3. Replace monitor.
Export halted: USB
drive full
[ 540 ]
1. Removable drive is not
connected properly to
USB(A) port.
2. Drive “write protect” is
locked.
3. Drive is incompatible or
defective.
4. Drive is full.
1. Check connection.
2. Verify “write protect”
switch on the drive is set to
“unlock” position.
3. Replace drive. Contact
your Aspect Technical
Service representative for a
list of compatible drives.
4. Replace drive.
Export halted: BISx
not detected
[ 541 ]
1. BISx disconnected.
2. Defective BISx cable.
3. Defective BISx.
4. Defective monitor.
1. Connect BISx. Verify all
cable connections.
2. Inspect/repair cable at
connector end.
3. Replace the BISx.
4. Replace monitor.
Export halted: USB
drive full
[ 544 ]
(See Message 540)
Export halted: USB
drive not detected
[ 545 ]
(See Message 539)
Replace PIC. Possible
PIC problem.
[ 548 ]
1. Defective PIC.
2. Defective Sensor
Simulator.
1. Test another PIC with
Sensor Simulator.
2. If failure persists, replace
simulator and retest.

Table of Contents