SECTION 6 DIAGNOSTICS AND TROUBLESHOOTING
______________________________________________________________________
6-10
Status Messages: Possible Causes: Corrective Actions:
Possible PIC problem.
Check cable using
Sensor Simulator
[ 549 ]
Defective PIC 1. Test PIC using Sensor
Simulator or Test Sensor.
2. Replace PIC.
Unrecoverable BISx
Error /
BISx initialization
error [ 1000-1999 ]
1. Poor connection
between BISx monitor
cable and monitor
2. Defective BISx.
3. Defective monitor.
1. Follow on-screen
instructions. If necessary,
power OFF and unplug
power cord to shut down
monitor completely. Then
plug in and re-start
monitor.
2. Replace the BISx.
3. Replace the monitor.
System Error/
Unrecoverable
Monitor Error
[ 2000-2999 ]
Hardware Timer Error
[ 2902 ]
A system error has
occurred. The monitor may
stop operating.
A system timer error has
occurred.
1. Follow on-screen
instructions (if any).
2. Turn the monitor off, then
on again.
3. Reset monitor.
4. Replace monitor.
Data Export Error
[ 3000-3999 ]
The data export was not
successful.
1. Removable drive is not
connected properly to
USB(A) port.
2. Drive “write protect” is
locked.
3. Drive is incompatible or
defective.
4. Drive is full.
1. Check connection.
2. Verify “write protect”
switch on the drive is set to
“unlock” position.
3. Replace drive. Contact
your Aspect Technical
Service representative for a
list of compatible drives.
4. Replace drive.
Unable to Update
Software
[ 4001- 4999 ]
1. Software has already
been updated to the
current revision.
2. A software error has
occurred.
1. Check Revision.
2. Restart monitor.
3. Reset monitor.
4. Replace drive.
Trend Memory Error
[ 5000-5999 ]
A memory error has
occurred.
1. Restart monitor.
2. Reset monitor.
3. Replace monitor.