2.2 Technical Key Features
Infrared Components
The infrared components consist of the following:
• Ceiling-mounted infrared and video camera system.
•
Wall-mounted touch screen monitor (standard).
• Control room workspace with monitor, keyboard and mouse.
• Compact-sized computer rack for computer, power supplies, electrical safety devices, control
electronics and excess cables (ExacTrac Computer Cabinet).
X-Ray Components
The X-ray components consist of the following:
• T
wo X-ray tubes recessed into the linac room floor.
• Two ceiling-mounted amorphous silicon detectors (ExacTrac Flat Panels).
• ExacTrac Console.
• Dual 65kW X-ray generator with integrated transformer unit.
Optional Components
Some optional components are:
• Ceiling-mounted T
ouch Screen Monitor for convenient handling (optional, standard is wall-
mount solution)
• Two X-ray Tubes mounted on-floor in the linac room (optional, standard is in-floor solution).
• Tilt alignments of targeting volume via ExacTrac Robotics.
Technical Key Features
14 Site Planning Manual Rev
. 1.0