5.5 Ceiling Mounted In-Room Distribution Box
General Information
572mm
Drilling pattern
In Room Distribution Box
Boreholes 6mm
272mm
Box
Top Cover
Service Area
70mm
Figure 32
The following tables provides information on the drilling pattern and required space for service for
the in-room distribution box:
Component Description
Dimensions 600 x 300 x 170 mm (length x width x depth)
Weight 8 kg (fully equipped)
Safety Factor Four (refer to local regulations)
Material Fire Resistance UL94 V-2
Fire Protection Measures According to local regulations
Bolts
You must supply four bolts for ceiling fixation,
according to local regulations.
Location
The in-room distribution box (1.3) should be in-
stalled so that the cable outlet (edge of the
box) is:
Maximum 4 m away from the black interface
plate of the celling mounted monitor arm
(1.2) or
Maximum 7 m away from the wall mount of
the wall mounted monitor arm (1.2).
Maximum 4.5 m away from the external blue-
tooth module (3.2).
If cables cannot be routed directly, plan the dis-
tance between the two related devices respec-
tively shorter.
CEILING AND WALL MOUNTS
Site Planning Manual Rev. 1.0 ExacTrac Ver. 6.x 67