Example Circuit X-Ray Warning Output to Building Installation
Exactrac cabinet
Building Installation
External warning light circuit
ET Electronics Rack
ET ILK Insert
K(-)
A(+)
K(-)
A(+)
MINUS
PLUS
EMO Input
XR Warning Output
12V DC Output
DOOR Input
EXT ILK Connector
4
5
8
3
1
2
6
7
(internal switch, closed
in case of x-ray active)
C
NO
Figure 40
ExacTrac Warning Lights