Network Connection
The network connection is required for:
Patient data import and export, depending on the setup even in real-time.
ADI connection to the linac (positioning, patient loading, treatment authorization, etc.
A stable, reliable and safe network environment/connection needs to be provided in order
to guarantee safe and effective use of the ExacTrac system.
Network Performance
The network speed shall be minimum 10 Mbit/s (Brainlab recommends 1 Gbit/s).
The answering time shall be less than 100 ms.
The network shall be fully switched.
Brainlab requests an external firewall. Network security is your responsibility or the responsibility
of a respective responsible organization.
Virus Scanner
Brainlab recommends the use of a state of the art virus scanner. You or a responsible organization
is responsible for installation and maintenance and must consider Brainlab’s recommendations.
For detailed information, refer to the Clinical User Guide.
Introduction
90 Site Planning Manual Rev
. 1.0
ExacTrac Ver. 6.x