5. If the active C-code (indicated by “*”) is the correct one for the battery being
serviced, go to step 2 otherwise continue to the next step.
6. Select another C-code.
Press or to scroll through the C-codes in the adapter.
Press ENTER when the correct C-code is displayed. Press ENTER again to confirm
the changes. The selected C-code is now active, and is indicated by “*”.
To change any of the settings in the C-code, do it now (see Create or Edit a C-
code, page 62).
Always use the correct C-code, especially the correct chemistry.
Failure to do so may damage the battery or cause fire or explosion.
1. If the C-code is correct, press ENTER to confirm the settings.
2. Press ENTER again to start the battery service. The yellow (run) light appears.
Once the service starts, the display will show the current program status as in the
following diagram. For detailed information about messages that appear, (see
Appendix A - Messages and Warnings, page 123).
Depending on the battery chemistry and the program selected, service may take
anywhere from a few seconds to a few hours. Some programs can take several
days or weeks.
Cadex Electronics Inc. C7000-C Series Battery Analyzers .