4.3.16. CAENVME_SetPulserConf
Parameters:
[in] Handle : The handle that identifies the device.
[in] PulSel : The pulser to configure (see CVPulserSelect enum).
[in] Period : The period of the pulse in time units.
[in] Width : The width of the pulse in time units.
[in] Unit : The time unit for the pulser configuration (see
CVTimeUnits enum).
[in] PulseNo : The number of pulses to generate (0 = infinite).
[in] Start : The source signal to start the pulse burst (see
CVIOSources enum).
[in] Reset : The source signal to stop the pulse burst (see
CVIOSources enum).
Returns:
An error code about the execution of the function.
Description:
The function permits to configure the pulsers. All the timing parameters
are expressed in the time units specified. The start signal source can be
one of: front panel button or software (cvManualSW), input signal 0
(cvInputSrc0),input signal 1 (cvInputSrc1) or input coincidence
(cvCoincidence). The reset signal source can be: front panel button or
software (cvManualSW) or, for pulser A the input signal 0 (cvInputSrc0),
for pulser B the input signal 1 (cvInputSrc1).
CAENVME_API
CAENVME_SetPulserConf(long Handle, CVPulserSelect PulSel, unsigned char Period,
unsigned char Width, CVTimeUnits Unit, unsigned char PulseNo, CVIOSources Start,
CVIOSources Reset);