EasyManuals Logo

CEM Phoenix User Manual

CEM Phoenix
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
99
Speci cations
Temperature Capabilities: 1000 °C ± 3 °C (Standard Furnace)
1200 °C ± 3 °C (High Temperature Furnace)
Electrical Ratings: 208/230 VAC, 60 Hz, 15 amps
220/240 VAC, 50 Hz, 15 amps
Voltage selection switch permits instrument to
operate properly with line voltages from
208/230 (60 Hz) or 220/240 (50 Hz)
Magnetron Frequency: 2450 MHz
Exhaust Capabilities: 100 CFM
Printer Port: 25-Pin Parallel (System compatible with
IBM, Epson, Okidata, ASCII, Star
NX-1020 and HP Thinkjet printers)
Computer Interface: RS-232 Serial Interface (PC software to
interface and collect data - must be supplied by
the user)
Balance Interface: RS-232, Serial Interface
Overall Instrument Dimensions: 18.17 x 25.75 x 19.60 in. (WxDxH)
46.15 x 65.40 x 49.78 cm (WxDxH)
Furnace Chamber Volume: 263.4 in
3
(5.0 liters) (Standard Furnace)
112.5 in
3
(1.8 liters) (High Temperature Furnace)
Internal Furnace Chamber Dimensions: Standard Furnace
9.7 x 6.6 x 4.0 in (WxDxH0
24.6 x 16.7 x 10.1 cm (WxDxH)
High Temperature Furnace
8.25 x 4.55 x 3.00 in. (WxDxH)
20.96 x 11.56 x 7.62 cm (WxDxH)
Unit Weight: 97.5 lbs.
44.226 kg.
IR Detection System: Infrared detection device for door protection
Internal Calibration of Calibrated to NIST Traceable Standard
Temperature Control:
Internal Diagnostic Software: BITS System (Built-in Test System)
Analyzes magnetron hrs., system power on hrs.,
time spent at various temperatures, number of
door openings and closings, calibration time
and date, calibration set points, calibration
tables and data
Safety Interlock: Three safety door interlock mechanisms prevent
instrument operation and microwave emissions in
case of improper door closure or misalignment