33
Quick Test - External Balance
Quick test permits entry of time and temperature parameters for a one-time ashing
procedure using either the standard or syringe weighing format. Ashing procedures
performed with the Quick Test method cannot be saved in the instrument memory.
To enter and store a method in the instrument directory, refer to the “Edit/Create
Method” section of this manual.
CEM Method Menu
EDIT/CREATE QUICK TEST LOAD
METHOD METHOD
Current Method: XXX
Press START to Begin
1. Using the arrow keys, highlight Quick Test. Press the SELECT key.
Method Setup
DWELL TIME: 00:00
OPERATING TEMP.; XXX
OPTION: Standard
2. Using the arrow keys, highlight Dwell Time. Use the numeric keys to enter
the desired method dwell time. Dwell time can be set from 1 minute to 99 hours
and 99 minutes. (Examples: To enter 5 minutes, press “5;” to enter 1 hour, press
“1, 0, 0.”)
3. Using the arrow keys, highlight Operating Temperature. Use the numeric keys
to enter the desired test temperature.
NOTE
The maximum operating temperature of the standard furnace is
1000 C. The maximum operating temperature of the high tempera-
ture furnace is 1200 C.
Note: If a value greater than 1200°C is entered, the Phoenix will automatically
default to 1200°C.
Note: The maximum operating temperature of the standard furnace is 1000°C, and
the maximum operating temperature of the high temperature furnace is 1200°C.
4. Using the arrow keys, highlight Option. Press the SELECT key to toggle and
select Standard or Syringe.