81
Maintenance, Troubleshooting and Service
Routine maintenance should be performed on the instrument to prolong the life of the
instrument and minimize downtime. Basic operational troubleshooting procedures
should be limited to replacement of defective sub-system components. For detailed
instructions concerning service and repair, contact the CEM Service Department or
the local subsidiary or distributor.
Routine Maintenance
A monthly routine preventive maintenance program is recommended to ensure opti-
mum performance of the Phoenix.
NOTE
Failure to properly maintain the Phoenix instrument will nullify the
instrument warranty.
Door, Seals and Interlocks – Carefully inspect the door, door seals, and door interlock
to verify that they are clean and working properly. Ensure that there is no loosening
of or damage to the door hinges or latch. Ensure that the door closes securely.
Cleaning Recommendations – Routine cleaning of the instrument is necessary to
prevent carbon buildup which can restrict air fl ow through the system, clog exhaust
ports, and render the thermocouple inoperative. Scheduled (monthly) cleaning of the
door seals, cavity interior and exhaust blower are necessary. For applications where
carbonaceous buildup is more prevalent, a weekly inspection should be performed
to detect the necessity for more frequent cleaning. Inspect and clean the Phoenix
as follows:
1. Permit the instrument interior surfaces to cool to room temperature. Unplug the
instrument from the electrical outlet.
2. Remove the screw in the thermocouple access door on top of the instrument
cover and remove the access door.
3. Loosen the thermocouple nut and slide the thermocouple out of the instrument
cavity. Using a soft cloth, clean the thermocouple. Inspect the thermocouple for
any pitting and/or wear. Replace the thermocouple as required. Place the
thermocouple in an area where it will be safe from damage.
4. Remove the muffl e furnace chamber.
5. Loosen the exhaust hose clamp and remove the exhaust hose.