34. Wait for the instrument to determine the nal weight and calculate the data
results.
READING BALANCE
Data Results
Sample ID: XXXX
Program: XX:XX at XXX C
Crucible Wt: X.XXXX
Start Wt: X.XXXX
Final Wt: X.XXXX
% XXX: XX.XX%
35. Press the PRINT key to print the data results or the “computer” key to send
the data to a PC. To continue weighing ashed samples, press the NEXT key.
36. To process additional samples using the same heating parameters, press the
HOME key.
37. To end the standard Quick Test method and return to the CEM Method Menu,
press the STOP key, followed by the HOME key.
Syringe
Note: To perform a Quick Test using the Syringe method, ensure that steps 1 through
15 have been performed.
SAMPLE ID: XXXX
CRUCIBLE WT: XXXX.X
Prepare Syringe with Sample,
Clear external balance pan
and press NEXT
38. Ensure that the external balance is empty.
39. Press the NEXT key.
Note: The sample and crucible must be cooled to ambient temperature prior to
reweighing.
32. Place the crucible with the ashed sample on the external balance pan.
33. Press the NEXT key.