76
28. Open the instrument door. Remove the furnace door and place it in the furnace
door holder.
29. Using metal tongs, remove the crucible(s) containing the sample(s) from the
ashing furnace.
30. Install the furnace door and close the instrument door. Press the START key.
31. Press the “Balance” key.
Edit Sample ID
ENTER NEW ID
FINAL WEIGHT
ERASE ID
SELECT ANALYSIS % ASH
32. Using the arrow keys, highlight Final Weight. Press the SELECT key.
Select Sample ID
XXXXX
XXX
XXXX
XXXX
XXX
33. Using the arrow keys, highlight the desired sample identifi cation. Press the
SELECT key.
FINAL WEIGHT
SAMPLE ID: XXXX
CRUCIBLE WT: XXXX.X
SAMPLE WT: XXXX
Clear the external balance pan
and press NEXT
34. Ensure that the external balance is clear. Press the NEXT key.
TARING BALANCE
35. Wait for the instrument to tare the balance.