Chison Q9 - 4.3.21. Power Doppler(CPA) Mode imaging optimization

Chison Q9
201 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Q9 Digital Color Doppler Ultrasound System
56
Blood Effection
Choose different blood effection
Smooth, HighRes
Shortcut Adjustment
Shortcut
action
Menu Menu Function Menu Item adjustable Rang
SK1
press
Frequency
Adjust the frequency of
probe
Refer to current probe
SK1
rotate
Steer
Adjust angle of sampling box
of blood flow under the linear
probe
Adjust range from -20 to 20
SK2
press
Wall Filter Change of Wall filtering 0~3 adjustable
SK2
rotate
Color Map Change of Color Map Adjust option:User, Type:1- 9
SK3
press
Flow Change sensitivity of blood High, Middle and Low for option
SK3
rotate
PRF Adjust PRF value
The maximum PRF depends on the probe and the
position of sampling box.
SK4
press
Invert Realize invert of blood flow Turn on or Turn of the invert
SK4
rotate
Persistence
Improve the current color 0~7 grade adjustable. Adjust range: 1.
SK5
press
Density Change images density High and Low adjustable.
SK5
rotate
Baseline
Change the baseline of color-
map
-3~3 adjustable.
Other adjustments
1 CFM gain control.
Rotate C- knob after C lights is on .Adjust the color gain and adjust range is from 0 to 255.
2 Sampling box.
Moves sampling box through trackball to the area which you are interested and press the ENTER- key to confirm the
position of sampling box then adjust the size of box trough trackball and press ENTER- key to confirm.
Use the STEER toggle key to change the direction of sampling box under the linear probe.