Normally open, Normally closed.
When programming the latch:
Enable the latch, Disable the latch.
The latch is selected in this parameter, i.e., if the relay digital output is locked after it has been
tripped, even when the condition that triggered it has disappeared.
Note: If the device loses its power supply, the latch disappears.
5�7�17�- TRANSISTOR DIGITAL OUTPUTS�
Select the output to configure on the main configuration screen of the transistor digital
outputs,Figure 295.
Figure 295: Main conguration screen of transistor digital outputs.