Codonics Virtua User’s Manual 11-19
Device Status and
Troubleshooting
7. Touch the
Calibrate
button.
Calibration status messages are displayed in the Dashboard. The
robotic disc transport will attempt to execute the operation,
according to the calibration type selected in step 4.
8. Watch the execution of the operation to verify whether the
alignment has been corrected. Repeat the calibration value
adjustment until the operation is performed successfully.
9. When the calibration completes successfully, touch the
Done
button to save the new calibration settings.
NOTE:
You can also touch the
Cancel
button at any time to discard the calibration setting
changes.