Codonics Virtua User’s Manual 3-29
Setting Up the Software
includedDicomViewer = Specifies the name of the file that contains the
DICOM viewer application to be included on the disc.
Choices are: Codonics_Viewer.zip (the default
viewer), Japanese_Viewer.zip. An eFilm Lite viewer
(eFilm_Lite.zip) can be purchased separately.
If the value is blank, no viewer will be included on the
disc.
In addition to eFilm Lite, some other third-party
viewers are supported. Contact Codonics for more
information about third-party viewers.
mediaType = Specifies the type of media on which to record the
studies.
Choices are: CD, DVD, AUTO.
AUTO causes the device to select the correct media
depending on the size of the job. Jobs requiring less
than 650 MB are recorded on CDs; those requiring
more than 650 MB are recorded on DVDs.
Jobs queued for a media not currently configured will
be marked as an error in the Studies list. For example,
a job queued for DVDs when both input bins are
configured for CDs will not record.
discStudyGrouping = With the doAutoRecord parameter set to True,
specifies whether the images received for recording
will be grouped for recording to disc by patient or
study.
Choices are:
byPatient. If multiple studies for the same patient
are received within the endOfStudyTimeout time,
the studies will be recorded on the same disc.
byStudy. One study per disc will be recorded.
Table 3-5. Job Profile Format and Parameters (Continued)